• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD EDS3 DRAINAGE SYSTEM; CSF DRAIN SYS/CRANIAL ACCESS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD EDS3 DRAINAGE SYSTEM; CSF DRAIN SYS/CRANIAL ACCESS Back to Search Results
Model Number 82-1730C
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported a broken thread between collection bag and eds3 drainage system.The issue happened when changing the collection bag and the drainage system had to be replaced on (b)(6) 2022 which took 20 minutes and patient care was delayed 1 hour.
 
Additional Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h6, h10.The eds3 drainage system was returned for evaluation: failure analysis - eds iii without collection bag was returned for investigation.Eds was visually inspected, the connector from the drip chamber to the collection bag was broken, stress marks were noted in the in the connector this is probably due to atypical force when connecting the collection bag.The eds was leak tested failed, leaked from the cracked connector.The current quality inspection for these assemblies is established to 100% test for leaks and blockage.The root cause of the problem reported by the customer is probably due to users error by using atypical force when attaching the collection bag, as noted in the ifu ¿finger tighten only when attaching devices¿.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDS3 DRAINAGE SYSTEM
Common Device Name
CSF DRAIN SYS/CRANIAL ACCESS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14565870
Report Number3013886523-2022-00230
Device Sequence Number5083009
Product Code JXG
UDI-Device Identifier10886704079866
UDI-Public10886704079866
Combination Product (Y/N)N
PMA/510(K) Number
K172537
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number82-1730C
Device Catalogue Number821730C
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/16/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date06/01/2022
Supplement Report FDA Received Date07/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-