| Model Number |
G24654 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Material Too Soft/Flexible (4007)
|
| Health Effect - Clinical Code |
Gastrointestinal Hemorrhage (4476)
|
| Date of Event |
05/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A corrective action has been initiated to reduce occurrences of the mbl bands not tight enough to function appropriately.The product said to be involved is included in the scope of the corrective actions.Prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a corrective action has been initiated in an effort to reduce occurrences of this nature.This product was manufactured prior to implementation of this corrective action.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During an endoscopic variceal ligation (evl), the physician used a cook 6 shooter saeed multi-band ligator.It was reported that the physician tried to deploy the band on the bleeding varices but the band slipped off [varices], so the physician opted to use a cook duette kit instead and successfully completed the procedure.An unintended section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Search Alerts/Recalls
|