Consumer reported complaint for high and erratic blood glucose test results.The customer is concerned with test results from results obtained of 511, 470, 337, 412 and 331 mg/dl.The customer¿s expected 2 hour post meal blood glucose test result range is 140-160 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The customer was not using the proper testing technique.During the call, a back to back blood test was performed by the customer fasting and produced test results of 317 mg/dl and 388 mg/dl using true focus meter.The product is stored according to specification in the office.The test strip lot manufacturer¿s expiration date is 03/31/2023 and open vial date is (b)(6) 2022.The meter memory was reviewed for previous test result history: (b)(6).
|