Model Number STRIP, TF WGN 50CTMG/DL 956970
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 05/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
Consumer reported complaint for high and erratic blood glucose test results.The customer is concerned with test results from results obtained of 511, 470, 337, 412 and 331 mg/dl.The customer¿s expected 2 hour post meal blood glucose test result range is 140-160 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The customer was not using the proper testing technique.During the call, a back to back blood test was performed by the customer fasting and produced test results of 317 mg/dl and 388 mg/dl using true focus meter.The product is stored according to specification in the office.The test strip lot manufacturer¿s expiration date is 03/31/2023 and open vial date is (b)(6) 2022.The meter memory was reviewed for previous test result history: (b)(6).
 
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Meter and test strips were not returned for evaluation.Manufacturer contacted customer in a follow-up call on (b)(6) 2022 to ensure the replacement products resolved the initial concern - able to establish contact with customer who stated replacement products resolved initial concern.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUE FOCUS
Common Device Name
SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key14566090
Report Number1000113657-2022-00309
Device Sequence Number5778988
Product Code NBW
UDI-Device Identifier00021292008956
UDI-Public(01)00021292008956
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K131013
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/31/2023
Device Model NumberSTRIP, TF WGN 50CTMG/DL 956970
Device Lot NumberNY1008
Was Device Available for Evaluation? No
Distributor Facility Aware Date05/04/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/04/2022
Initial Report FDA Received Date06/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-