BAXTER HEALTHCARE CORPORATION EXTENSION SETS WITH ONE-LINK NEEDLE-FREE IV CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 7N8301 |
Device Problem
Incomplete or Inadequate Connection (4037)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/03/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the tubing of an one-link non-dehp standard bore catheter extension set had a defect in the plastic.Additionally, "the male end of the extension set that should go into the female of the ll on the iv tubing had a defect in the plastic which made it impossible to attach".Llk, lp, l; the issue was identified during setup and preparation.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
H4: the lot was manufactured from november 14, 2021 - november 15, 2021.H10: the device was received for evaluation.Visual inspection was performed using the naked eye which observed that the end of the extension set has a defect in the plastic of the male luer.The reported condition was verified.The cause of the condition was due to a manufacturing related issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|