• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EXTENSION SETS WITH ONE-LINK NEEDLE-FREE IV CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EXTENSION SETS WITH ONE-LINK NEEDLE-FREE IV CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 7N8301
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/03/2022
Event Type  malfunction  
Manufacturer Narrative
Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing of an one-link non-dehp standard bore catheter extension set had a defect in the plastic.Additionally, "the male end of the extension set that should go into the female of the ll on the iv tubing had a defect in the plastic which made it impossible to attach".Llk, lp, l; the issue was identified during setup and preparation.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured from november 14, 2021 - november 15, 2021.H10: the device was received for evaluation.Visual inspection was performed using the naked eye which observed that the end of the extension set has a defect in the plastic of the male luer.The reported condition was verified.The cause of the condition was due to a manufacturing related issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SETS WITH ONE-LINK NEEDLE-FREE IV CONNECTOR
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14566704
MDR Text Key293096926
Report Number1416980-2022-02802
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412475387
UDI-Public(01)00085412475387
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132734
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number7N8301
Device Lot NumberR21K13017
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/05/2022
Initial Date FDA Received06/01/2022
Supplement Dates Manufacturer Received07/04/2022
Supplement Dates FDA Received07/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-