| Model Number |
70914-70 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product has been returned, extended investigation is pending at this time.A follow- up report will be submitted once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving a low reading from their adc device.Customer reported a low reading of 120 mg/dl which was lower than lab reading of 264 mg/dl.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs (device history review) for the freestyle freedom lite meter were reviewed and the dhrs showed the freestyle freedom lite meter passed all tests prior to release.Dhrs (device history review) for the freestyle strips were reviewed and the dhrs showed the freestyle strips passed all tests prior to release.Retain testing was performed for freestyle strips and all units performed in specification and passed.Meter (b)(6) has been returned and investigated with retained test strips.Visual inspection has been performed on the meter and no issues were observed.Meter powered on with button depression and with strip insertion.Control solution testing was performed and no issues were performed.All results were within range specification.No malfunction or product deficiency was identified.If the reported test strips are returned, an investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving a low reading from their adc device.Customer reported a low reading of 120 mg/dl which was lower than lab reading of 264 mg/dl.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|