• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL INC. LTV 1150 VENTILATOR; VENTILATOR,CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL INC. LTV 1150 VENTILATOR; VENTILATOR,CONTINUOUS, FACILITY USE Back to Search Results
Model Number LTV 1150
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect device was returned and evaluation is anticipated but not yet begun.Once a final investigation is complete, a follow-up report will be submitted.
 
Event or Problem Description
The customer reported problem with lap top ventilator 1150 where internal battery failed causing unit to power down while transporting a patient from er to icu.No patient harm reported.A staff immediately noted the condition and began bagging the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LTV 1150 VENTILATOR
Common Device Name
VENTILATOR,CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL INC.
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key14566854
Report Number2021710-2022-16118
Device Sequence Number17379215
Product Code CBK
UDI-Device Identifier10190752115644
UDI-Public(01)10190752115644(11)20200801
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K060647
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLTV 1150
Device Catalogue Number18984-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2022
Initial Report FDA Received Date06/01/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/01/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-