| Model Number |
LTV 1150 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device was returned and evaluation is anticipated but not yet begun.Once a final investigation is complete, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported problem with lap top ventilator 1150 where internal battery failed causing unit to power down while transporting a patient from er to icu.No patient harm reported.A staff immediately noted the condition and began bagging the patient.
|
| |
|
Search Alerts/Recalls
|