Model Number 1884016HR
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care provider (hcp) reported that the blade did not work correctly during sinus surgery and had the staff save it.He proceeded with a different blade, but the staff did not know the details of what malfunctioned.On additional information, it indicated the interior of the blade became disconnected.There was no case delay no patient impact.No broken pieces of the reported product remain inside the patient's body.Additional information received and it clarified that connection issue was not an electrical connection issue, it was like the blades 2 pieces came apart and would not oscillate correctly.
 
Additional Manufacturer Narrative
H3: product analysis: visually, there was biological contaminant coating most of the device.The spiral wrap of the middle tube was expanded approximately 0.10 inches from the distal end of the outer sheath which would have resulted in the reported event.There was no damage to the distal tip.There were no loose components.Functionally, there was binding while rotating the inner assembly by hand, but the middle assembly could not rotate at all.Console functional testing was not performed due to the aforementioned defects.There was no allegation of a defect prior to use.A review of the global complaint data showed three similar complaints about this lot number.Conclusion: in the returned condition, there was an out of specification condition that was related to the complaint (due to physical damage).Earlier submitted code b21, c21 and d16 are no longer applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BLADE
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key14566903
Report Number1045254-2022-00295
Device Sequence Number5231867
Product Code EQJ
UDI-Device Identifier00885074080642
UDI-Public00885074080642
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/02/2023
Device Model Number1884016HR
Device Catalogue Number1884016HR
Device Lot Number0217636379
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2022
Supplement Date Received by Manufacturer06/07/2022
Initial Report FDA Received Date06/01/2022
Supplement Report FDA Received Date07/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/03/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age32 YR
Patient SexMale
-
-