| Model Number |
8884711006E |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that during assessment and flushing of the dobhoff tube, the dobhoff tube cracked at top.The dobhoff tube had to be removed and a new one placed.Additional information received on (b)(6) 2022 stated that the damage occurred at the point where the plastic port fits into the rubber y which caused leakage when formula and flushes were administered.It is unknown how long the feeding tube was in place.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record review could not be performed because the lot number provided by the customer is not a valid lot number.The device was received for evaluation and the reported condition has been confirmed.A corrective and preventive action has been initiated to address the reported issue.
|
| |
|