Model Number N/A
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refw2775 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "nurse telling me there is no re- recanulate when she met resistant she removed immediately, so there was no opportunity for that to happen.This is the 28 cm device.".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of powerglide placement complications was confirmed.The product returned for evaluation was one 20ga x 8cm powerglide pro rt midline catheter assembly.Usage residues were observe throughout the sample.The sample was received assembled with the catheter overlaying the needle shaft.The needle protruded from a split in the catheter approximately 2cm from the distal tip.The guidewire remained inlaid through the catheter.Microscopic inspection of the split in the catheter revealed a glossy fracture surface.Inspection of the distal end of the catheter revealed deformation throughout the catheter tip.The fracture features were consistent with catheter perforation by the needle tip.Such damage can occur if the catheter is pulled back onto the needle and if the needle is re-inserted following catheter advancement.The catheter tip deformation was consistent with attempted catheter advancement against resistance, such as into tissue.H3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported "nurse telling me there is no re- recanulate when she met resistant she removed immediately, so there was no opportunity for that to happen.This is the 28 cm device.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL KIT 20G X 8 CM WITH GUARDIVA AND PROBE COVER
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14568495
Report Number3006260740-2022-02032
Device Sequence Number17663543
Product Code FOZ
UDI-Device Identifier00801741140570
UDI-Public(01)00801741140570
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K162377
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberF120088PT
Device Lot NumberREFW2775
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/10/2022
Supplement Date Received by Manufacturer06/24/2022
Initial Report FDA Received Date06/01/2022
Supplement Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-