• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, 1588 AIM CAMERA CONTROL UNIT (CCU); CONFOCAL OPTICAL IMAGING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, 1588 AIM CAMERA CONTROL UNIT (CCU); CONFOCAL OPTICAL IMAGING Back to Search Results
Model Number 1588010000
Medical Device Problem Code Electrostatic Discharge (2149)
Health Effect - Clinical Codes Electric Shock (2554); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was electrical shock to the user.The user was not injured.
 
Additional Manufacturer Narrative
The product was returned for investigation and the reported failure mode was not confirmed.Alleged failure: the primary lost image during the procedure for a few minutes.Root cause: it is difficult to determine the root cause of the issue observed at the account since we are unable to replicate the problem in-house.Possible root causes could be loose power cables or other unit such as an electrosurgical generator or cautery pencil near the ccu.The failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that there was electrical shock to the user.The user was not injured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 1588 AIM CAMERA CONTROL UNIT (CCU)
Common Device Name
CONFOCAL OPTICAL IMAGING
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key14569265
Report Number0002936485-2022-00282
Device Sequence Number17325478
Product Code OWN
UDI-Device Identifier07613327174328
UDI-Public07613327174328
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K142310
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1588010000
Device Catalogue Number1588010000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/05/2022
Supplement Date Received by Manufacturer05/05/2022
Initial Report FDA Received Date06/01/2022
Supplement Report FDA Received Date07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1