• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 10MM X 40CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 10MM X 40CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0036121040
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The subject device is unavailable to manufacturer.
 
Event or Problem Description
It was reported that during posterior communicating artery (pcom) aneurysm embolization procedure, the subject coil was delivered via the microcatheter into the aneurysm.At some point, the subject coil was not able to advance and not fully inserted into the aneurysm.The next step was to withdraw the whole subject coil into the microcatheter.Upon withdrawn, the subject coil was stretched and was noted to be broken into two parts.Most of the subject coil was withdrawn with the microcatheter but the broken/stretched distal part of the subject coil remained inside the patient¿s vessel.It was observed that the thin fiber rolled to a "ball of fibers" and this ball was removed successfully with a snare device.Patient was set under ass and is stable.
 
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes) system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not returned for analysis; therefore, physical as well as a functional testing could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The device was not returned and therefore the as reported cannot be confirmed.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable will be assigned to this complaint.
 
Event or Problem Description
It was reported that during posterior communicating artery (pcom) aneurysm embolization procedure, the subject coil was delivered via the microcatheter into the aneurysm.At some point, the subject coil was not able to advance and not fully inserted into the aneurysm.The next step was to withdraw the whole subject coil into the microcatheter.Upon withdrawn, the subject coil was stretched and was noted to be broken into two parts.Most of the subject coil was withdrawn with the microcatheter but the broken/stretched distal part of the subject coil remained inside the patient¿s vessel.It was observed that the thin fiber rolled to a "ball of fibers" and this ball was removed successfully with a snare device.Patient was set under ass and is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 10MM X 40CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14569647
Report Number3008881809-2022-00259
Device Sequence Number6085042
Product Code HCG
UDI-Device Identifier07613327004601
UDI-Public07613327004601
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/07/2023
Device Model NumberM0036121040
Device Catalogue NumberM0036121040
Device Lot Number22474541
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer08/12/2022
Initial Report FDA Received Date06/01/2022
Supplement Report FDA Received Date08/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/08/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SL-10 MICROCATHETER (STRYKER); TARGET COILS (STRYKER)
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-