MERIT MEDICAL MANUFACTURING IMPRESS®.BRAIDED CATHETER; PERIPHERAL CATHETERS Back to Search Results
Catalog Number 56538KA2
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The account alleges that during an endovascular procedure, the catheter tip detached within the patient.The physician had acquired vascular access and during catheter manipulations over a guidewire under fluoroscopy the catheter tip detached.The physician was able to successfully externalized the foreign body from the patient with a vascular snare device thus liberating the vessel.The patient tolerated the procedure well.No additional patient consequences to report.
 
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
The suspect device was returned for investigation.The device was examined visually and microscopically.The complaint is confirmed.The root cause could not be determined.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS®.BRAIDED CATHETER
Common Device Name
PERIPHERAL CATHETERS
Manufacturer (Section D)
MERIT MEDICAL MANUFACTURING
14646 kirby dr
houston TX 77047
Manufacturer (Section G)
MERIT MEDICAL MANUFACTURING
14646 kirby dr
houston TX 77047
Manufacturer Contact
bryson heaton bsn,rn.
1600 merit parkway
8012084662
MDR Report Key14569837
Report Number3010665433-2022-00053
Device Sequence Number20377768
Product Code DQO
UDI-Device Identifier00884450009437
UDI-Public884450009437
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K053171
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/17/2025
Device Catalogue Number56538KA2
Device Lot NumberE2354769
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer08/23/2022
Initial Report FDA Received Date06/01/2022
Supplement Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDWIRE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-