We checked the returned unit and confirmed that the angle wire was broken.Based on the result, we concluded that it was caused due to the excessive force applied on the angle wire.In addition, we confirmed that the angulation-down angulation zero degrees; however, it is not the main cause, and/or irrelevant to the alleged complaint.As a result of confirming the detail as gfe, no response was obtained, so we cannot deny the possibility of the angle stuck.Therefore, based on the technical report ""hr-rpt-0587(angle)"" and/or the risk analysis results, it was evaluated to submit mdr.
|