Model Number CQF7588
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one conquest 40 balloon dilation catheter was returned for evaluation.Balloon rupture was noted on the returned device.No other specific anomalies noted during the visual evaluation.In the functional testing, the balloon was inflated with the in-house presto inflation device and water was leaking from the balloon; the balloon fibers were stripped, and under microscopic examination a compound rupture was noted to the balloon.No other functional testing was performed.All the anomalies noted on the microscopic observation.Therefore, the investigation for the reported balloon rupture was confirmed as a compound rupture was noted during the microscopic observation and balloon leaks from the rupture during the functional testing of the returned device.A definitive root cause for the reported balloon rupture could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 09/2024).
 
Event or Problem Description
It was reported that during an angioplasty procedure, the balloon was allegedly burst.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST 40
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14572945
Report Number2020394-2022-00434
Device Sequence Number12022310
Product Code DQY
UDI-Device Identifier00801741060571
UDI-Public(01)00801741060571
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K120660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQF7588
Device Catalogue NumberCQF7588
Device Lot NumberREFX4157
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2022
Initial Report FDA Received Date06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-