| Model Number |
NSEAL535RH |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent the lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.
|
| |
|
Event or Problem Description
|
|
It was reported that during a lap colectomy, surgeon was using the enseal during the colon resection portion when the tip of the device broke off.Surgeon noticed device failure and retrieved the tip portion of the device.Device was removed.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 6/20/2022.D4 batch #: x94g17.Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the nseal535rh device was returned with the upper jaw detached and not returned.Additionally, the top of the iblade was broken off and not returned.The device was connected to the generator and it was recognized.Because the jaw was detached not all functional testing could be performed with the generator.A probable cause of the damage could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.As described in the ifu: do not use excessive force on the closing handle to close the jaws; grasp only as much tissue as will fit between the jaws where the current will pass.Greater amounts of tissue require more closing handle force.Excessive force could damage the device.Please reference the instruction for use for more information: as part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|