According to the reporter, before placing the device in the patient, the doctor noticed that the cuff of the device did not have the natural conical or oval shape that these devices generally had and that an unnatural swelling was observed on one of its sides.There was no patient involvement.
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted the overall shape of the cuff was not asymmetrical.It was reported that the cuff was distorted and not symmetrical.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: pre-insertion cuff and inflation test.With shiley cuffed models (lpc, fen) the cuff and inflation system should be tested for leakage before inserting the tube.This test can be performed as follows: inflate the cuff with the volume of air indicated in table 1.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Additional information: d9, g1(manufacturer name, mfr contact first name, last name, street 1, mfr city, region, postal code, email and phone number), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the cuff was distorted in one side of the cuff.Functionally, a dimensional test was performed.Single wall thickness was measured and the result was within specification.It was reported that the cuff was distorted and not symmetrical.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: with shiley cuffed models (lpc, fen) the cuff and inflation system should be tested for leakage before inserting the tube.This test can be performed as follows: inflate the cuff with the volume of air indicated in table 1.Then either observe for deflation over several minutes or immerse the tube in sterile saline and observe for air leakage.Deflate the cuff prior to insertion.Under no circumstances should more than 25mm of mercury air pressure be used to inflate the cuff.Over-inflation of the cuff may cause tracheal damage and may inhibit ventilation.To ease insertion and to guard against cuff perforation from sharp edges of cartilage, the cuff should be tapered back.This can be accomplished by first inflating the cuff.Then gently move the cuff away from the distal tip of the outer cannula towards the swivel neck plate as the residual air is removed by deflation.Do not use any sharp instruments such as forceps or hemostats that would damage the cuff when tapering it.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|