MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTERLOCK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTERLOCK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83779
Medical Device Problem Codes Entrapment of Device (1212); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the coil protruded from the catheter's tip during removal.The non-stenosed target lesion was located in the severely tortuous and non-calcified splenic artery.A 22mm x 60cm interlock coil was selected for use.During the procedure, the coil became stuck in the distal part of the catheter.When the coil was removed from the patient's body, it was noted that the coil protruded from the tip of the microcatheter.Only the coil was removed and the procedure was completed with another of same device.No patient complications were reported.
 
Additional Manufacturer Narrative
(b)(6).
 
Additional Manufacturer Narrative
E1 - initial reporter address 1: (b)(6).Device evaluated by mfr: the device was returned for evaluation.The device returned with a main coil, pusher wire, and introducer sheath.Visual inspection was performed, and no damages were found in the introducer sheath; however, it was returned with blood showing evidence of use.The twist lock was opened.The main coil and the pusher wire were interlocked inside the introducer sheath.The main coil and introducer sheath were inspected, and no anomalies or damages were observed.The pusher wire was found kinked.No more damages were observed.Functional inspection was performed, and the pusher wire and the coil were advanced through the introducer sheath.It was pulled out backwards in order to release the main coil without issues.Microscope inspection of the main coil and the pusher wire revealed that the proximal end has a smooth surface.The interlocking arm was inspected, and no anomalies was noted.Dimensional inspection of the main coil and the pusher wire revealed that the components were within specification.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that the coil protruded from the catheter's tip during removal.The non-stenosed target lesion was located in the severely tortuous and non-calcified splenic artery.A 22mm x 60cm interlock coil was selected for use.During the procedure, the coil became stuck in the distal part of the catheter.When the coil was removed from the patient's body, it was noted that the coil protruded from the tip of the microcatheter.Only the coil was removed and the procedure was completed with another of same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14575444
Report Number2134265-2022-06038
Device Sequence Number6147045
Product Code KRD
UDI-Device Identifier08714729845324
UDI-Public08714729845324
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K132578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/27/2024
Device Model Number83779
Device Catalogue Number83779
Device Lot Number0027919147
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/10/2022
Supplement Date Received by Manufacturer06/15/2022
Initial Report FDA Received Date06/02/2022
Supplement Report FDA Received Date07/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CATHETER - CARRY LEON LUMEN 0.022 INCH; CATHETER - CARRY LEON LUMEN 0.022 INCH
Patient SexUnknown
-
-