• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PEEL-AWAY INTRODUCER 14 CM SHEATH INTRODUCER KIT, 8F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PEEL-AWAY INTRODUCER 14 CM SHEATH INTRODUCER KIT, 8F; INTRODUCER, CATHETER Back to Search Results
Model Number 405112
Medical Device Problem Codes Fracture (1260); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
During implantation of a single chamber pacemaker, the sheath broke during access via the right subclavian artery.
 
Additional Manufacturer Narrative
Three images were submitted for evaluation to product performance engineering; no device components were returned.The photos appeared to show a peel-away sheath and dilator.The sheath appeared to have been punctured and partially split.Visual inspection was based solely upon a review of the photographs provided.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported sheath break remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEEL-AWAY INTRODUCER 14 CM SHEATH INTRODUCER KIT, 8F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL (
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key14575450
Report Number3005334138-2022-00353
Device Sequence Number6147533
Product Code DYB
UDI-Device Identifier05414734204125
UDI-Public05414734204125
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K894431
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model Number405112
Device Catalogue Number405112
Device Lot Number8174897
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer07/26/2022
Initial Report FDA Received Date06/02/2022
Supplement Report FDA Received Date07/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-