• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR-CALIF AXS VECTA 074 CATH 132CM - US; CATHETER, THROMBUS RETRIEVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR-CALIF AXS VECTA 074 CATH 132CM - US; CATHETER, THROMBUS RETRIEVER Back to Search Results
Model Number INC-11597-132
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
The device was returned for analysis and the investigation of the device revealed that the catheter (subject device) shaft was fractured during use on the patient.No clinical consequences were reported to the patient due to this event.No other information is available.
 
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The device was returned and the lot number was confirmed with the packaging returned with the device.During visual inspection the catheter was seen to be broken/fractured 100cm from catheter hub and was kinked under the strain relief.The hub id looked intact.During functional inspection resistance noted at kink under strain relief.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint was confirmed based on analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.The catheter was seen to be broken/fractured and was also kinked/bent.This complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.Therefor the analyzed event of the catheter shaft broken/fractured will be assigned procedural factors.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXS VECTA 074 CATH 132CM - US
Common Device Name
CATHETER, THROMBUS RETRIEVER
Manufacturer (Section D)
STRYKER NEUROVASCULAR-CALIF
47900 bayside parkway
fremont CA 94538
Manufacturer (Section G)
STRYKER NEUROVASCULAR-CALIF
47900 bayside parkway
fremont CA 94538
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14575471
Report Number3008853977-2022-00031
Device Sequence Number20375675
Product Code NRY
UDI-Device Identifier07613327459265
UDI-Public07613327459265
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K172167
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/10/2024
Device Model NumberINC-11597-132
Device Catalogue NumberINC-11597-132
Device Lot Number16978-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/11/2022
Initial Report FDA Received Date06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age80 YR
Patient SexFemale
-
-