|
|
| Model Number |
48003030X |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the manufacturing documentation associated with lot 82226771 presented no issues during the manufacturing process that can be related to the reported event.Additional information is pending and will be submitted within 30 days upon receipt.
|
| |
|
Event or Problem Description
|
|
As reported, there was a leakage coming from the balloon of a 3mm x 30cm 150cm saber percutaneous transluminal angioplasty (pta) balloon catheter which was unable to be fully inflated during an attempted inflation to 15 atmospheres (atm).A second 3mm x 30cm 150cm saber pta balloon catheter was used for the procedure and experienced the same issue of leakage, and an inability to be fully inflated.As a result, a third 3mm x 30cm saber pta balloon catheter was used to complete the procedure without issue.There was no reported injury to the patient.This was during a procedure to treat a lesion below the knee which had a 78% stenosis.The lesion was not calcified, tortuous, and was not a chronic total occlusion (cto).The initial 3mm x 30cm 150cm saber pta balloon catheter was stored and prepped per the instructions for use (ifu) and maintained negative pressure during preparation.A 1:1 contrast to saline ratio was used to prep the balloon.There was no difficulty removing the device from the hoop, removing the protective balloon cover, the stylet, or any of the sterile packaging components and the device was not kinked or damaged in any way prior to being inserted into the patient.There was no resistance or friction experienced during insertion of the initial 3mm x 30cm 150cm saber pta balloon catheter.The device was able to track towards the lesion and cross the lesion without issue.The initial 3mm x 30cm 150cm saber pta balloon catheter was never put into an acute bend and did not kink during its use.After the leakage was observed, the initial 3mm x 30cm 150cm saber pta balloon catheter was able to be removed easily from the patient and remained in one piece during its removal.Information regarding the preparation, insertion, and removal of the second 3mm x 30cm 150cm saber pta balloon catheter is still pending.The devices were discarded and will not be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
This is one of two products involved with the reported event and the associated manufacturer report number is: 9616099-2022-05689 after further review of additional information received the following sections have been updated accordingly: complaint conclusion: as reported, there was a leakage coming from the balloon of a 3mm x 30cm 150cm saber percutaneous transluminal angioplasty (pta) balloon catheter which was unable to be fully inflated during an attempted inflation to 15 atmospheres (atm).A second 3mm x 30cm 150cm saber pta balloon catheter was used for the procedure and experienced the same issue of leakage, and an inability to be fully inflated.As a result, a third 3mm x 30cm saber pta balloon catheter was used to complete the procedure without issue.There was no reported injury to the patient.This was during a procedure to treat a lesion below the knee which had a 78% stenosis.The lesion was not calcified, tortuous, and was not a chronic total occlusion (cto).The initial 3mm x 30cm 150cm saber pta balloon catheter was stored and prepped per the instructions for use (ifu) and maintained negative pressure during preparation.A 1:1 contrast to saline ratio was used to prep the balloon.There was no difficulty removing the device from the hoop, removing the protective balloon cover, the stylet, or any of the sterile packaging components and the device was not kinked or damaged in any way prior to being inserted into the patient.There was no resistance or friction experienced during insertion of the initial 3mm x 30cm 150cm saber pta balloon catheter.The device was able to track towards the lesion and cross the lesion without issue.The initial 3mm x 30cm 150cm saber pta balloon catheter was never put into an acute bend and did not kink during its use.After the leakage was observed, the initial 3mm x 30cm 150cm saber pta balloon catheter was able to be removed easily from the patient and remained in one piece during its removal.Information regarding the preparation, insertion, and removal of the second 3mm x 30cm 150cm saber pta balloon catheter is still pending.The devices were not returned for analysis as they were discarded.2022-00185282-1 a product history record (phr) review of lot 82226771 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.2022-00185282-2 a product history record (phr) review of lot 82226771 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon leakage - during positive pressure¿ could not be confirmed as the devices were not returned for analysis.The exact cause could not be determined.It is likely vessel characteristics of stenosis and/or procedural factors may have contributed to the reported events.However, without the return of the devices for analysis it is difficult to draw a clinical conclusion between the devices and the events reported.According to the instructions for use, which are not intended to mitigate risk, ¿preparation 1.Attach a 3-way stopcock to the inflation port, which is marked ¿balloon¿.2.Attach a partially filled syringe with heparinized saline to the stopcock, open the stopcock to the balloon and induce negative pressure.3.Hold the syringe and proximal end of the catheter above the distal end of the catheter, and hold the balloon vertically with the balloon tip pointing down.4.While maintaining negative pressure close the stopcock to the inflation port.Remove the syringe and purge the air.5.To ensure air contained in the balloon and inflation lumen is removed, apply negative pressure twice as instructed and repeat steps 2-4.6.Without twisting, slide the forming tube off the balloon.7.Prepare an angioplasty inflation system with a 50% solution of contrast medium in sterile saline or similar solution.8.Purge the air from the inflation device.9.Connect the inflation device to the 3-way stopcock that is connected to the catheter inflation port, open the stopcock to the catheter and slowly fill the inflation lumen and the balloon will slowly fill with diluted contrast medium.¿ neither the phr review nor the analysis results suggest that the reported event is related to the manufacturing process.Therefore, no corrective or preventive actions will be taken at this time.
|
| |
|
Search Alerts/Recalls
|
|
|