It was reported that fifteen minutes post placement of a stent, occlusive flow and a clot in the stent were discovered.Medication was administered resulting in a fully opened stent with full patency of the vessel.The treatment was for an aneurysm in the left vertebral artery.The stent was deployed without issue.Post deployment injections showed an open stent and full patency of the vessel.The anatomy was tortuous.Fifteen minutes after the procedure, the patient showed symptoms and an angiogram was performed showing occlusive flow and a clot in the stent.The physician went up with a microcatheter and administered integrilin at the proximal end of stent.The physician then proceeded to cross the stent with a j-shaped wire and was able to fully cross with a competitor catheter.The physician then proceeded to inject more integrilin as the physician pulled back on the catheter.Final angio was performed and indicated a fully open stent with full patency of the vessel.The patient was left intubated after rescue procedure and sent to icu.It was reported that the patient was a little better.The patient is alert and following some commands.The patient has left side neglect and weakness.Dvts in le's were found.It seems likely the patient has some kind of hypercoagulable state.Mri and ivc filter are pending.
|
|
In place of a physical product evaluation as series of images were provided and reviewed.There are a series of left vertebral subtracted dsas which demonstrate what is likely an atherosclerotic aneurysm of the distal (v4) segment of the left vertebral artery, close to the vertebra-basilar junction.The aneurysm is wide-necked and measures about 5x8 mm (my own estimate - no measurements are provided).Parent artery measurements are provided: proximal: 3.39mm, 3.86mm; distal: 4.27mm, 4.17mm; middle: 2.71mm.There are mild to moderate atherosclerotic irregularities of the left vertebral and basilar arteries.There is another small atherosclerotic aneurysm of the right distal basilar between the aica and the sca; it is broad-based with a neck of about 6mm and a depth of about 4mm.The right vertebral artery is likely occluded as there is retrograde flow into it from the left vertebral, down to the right pica, with no inflow of unopacified blood.Then, images (subtracted and unsubtracted dsas, with and without contrast) show a fully deployed fred with good apposition (on the images provided) spanning from proximal to the left pica to the very distal part of the left v4 vertebral artery; the most distal part of the fred is in a fairly tight curve.There is a very short segment of minimal spasm just proximal to the proximal radiopaque markers of the fred.A distal access catheter is located at the v3 segment.Subsequent images (subtracted and unsubtracted dsas, with and without contrast) show the development of clot in the distal part of the stent at first, and then in its proximal part.Ultimately, the whole stent thromboses with no antegrade flow.Subsequent images show a microcatheter navigated through the stent and distally to the distal basilar.Later images show the distal access catheter navigated to the distal part of the fred.Later images show progressive disappearance of the clot, with ultimately the fred being fully open, with no evidence of clot.There are no obvious distal emboli, but a full head ap and lateral completion angiogram of the left vertebral artery is not supplied.The cause for the clotting is not obvious in these images; the acute curve at the end of the device may have been a contributor, but it sounds from the report that the patient may have been hypercoagulable given the development of bilateral dvts.Without the return an physical evaluation of the device, the investigation cannot determine if a condition existed that would have caused or contributed to the reported event.
|