• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIGTRS45AMT
Medical Device Problem Codes Entrapment of Device (1212); Device Displays Incorrect Message (2591); Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a thoracoscopic lobectomy, while doing an interlobar formation, the initial reload did not fire.The second reload used did not fire as well.It was also noted that the jaws of the device locked on tissue and an exclamation mark appeared on the display.The device was removed and a third reload was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
D10 concomitant product: sigpshell signia (lot number: unknown), sigadaptstnd signia (serial number: unknown), sigphandle signia (serial number: unknown).Additional information: b5, d10, g1, g2, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the reload was partially fired and the interlock was engaged.The sled and staples were visible at 2.5cm cut lines.The jaw of the reload was open and the staple pushers were visible at the 3.1cm cut line.The proximal sutures were cut properly and the distal sutures were returned intact.The reinforcement materials were cut at the 2cm cut line.It was reported that the device did not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.It was also reported that the screen displayed a yellow swim lane aligned with the reload symbol.This indicates a general reload error.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: failure to completely fire the reload will result in an incomplete cut and/or incomplete staple formation, which may result in poor hemostasis and/or leakage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a thoracoscopic lobectomy, while doing an interlobar formation, the initial reload did not fire.The second reload used did not fire as well.It was also noted that the jaws of the device locked on tissue and an exclamation mark appeared on the display.The jaws were able to be opened and the device was removed from the tissue.A third reload was used to resolve the issue.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key14577695
Report Number1219930-2022-02079
Device Sequence Number6177322
Product Code GDW
UDI-Device Identifier10884521543645
UDI-Public10884521543645
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160176
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model NumberSIGTRS45AMT
Device Catalogue NumberSIGTRS45AMT
Device Lot NumberN1F0926Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer08/23/2022
Initial Report FDA Received Date06/02/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SIGTRS45AMT TRI-STAPLE 2.0 (LOT NUMBER: N1F0926Y).
Patient SexUnknown
-
-