| Model Number |
4251601-020 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun internal report number (b)(4).The complaint is still under investigation.A follow-up report will be provided, as soon as investigation has been completed.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility (translation of user facility information by bbm sales organization in (b)(6)."product leakage." "during priming of infusion preparation in the chemotherapeutic room, a liquid leak occurred from the jms infusion set connection.This is the second case at this department, and lot is different from the last time.About 5 same cases have occurred at this facility.The patient has no infection / health hazard to patient: none / sample: available, will be shipped later./ batch: 21h29g8371.".
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun internal report number (b)(4).Device history record (dhr): reviewed the device history record for batch number 21h29g8371 and there were no defects encountered during in process and final control inspection.Sample evaluation: received 1 used and contaminated introcan safety 3 pur 22g 0.9x25mm-ap in an opened packaging.Lot number printed on returned peel pack is 21h29g8371 and material# 4251128-03.Visual inspection: visual inspection from the 1 pc returned sample, no damage or dent at catheter luer.Air tightness test: sample was subjected to air tightness test at 300 kpa / 30 sec under water.Result: no leakage observed.Conclusion: no damages and no leakage were observed on the returned sample.Complaint is not confirmed.
|
| |
|
Search Alerts/Recalls
|