Model Number 7010
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that labeling issue occurred.A 2.5mm x 12mm quantum maverick balloon catheter was selected for use.However, during unpacking, the outer package, inner package and inner shaft of the balloon, did not match the model.The size on the outer package was 2.5mm x 12mm, and the size of the inner package is 2.75mm x 12mm, but the size on the delivery shaft is still 2.5mm x 12mm.Another quantum maverick balloon catheter was used to complete the procedure.There were no patient complications reported.
 
Event or Problem Description
It was reported that labeling issue occurred.A 2.5mm x 12mm quantum maverick balloon catheter was selected for use.However, during unpacking, the outer package, inner package and inner shaft of the balloon, did not match the model.The size on the outer package was 2.5mm x 12mm, and the size of the inner package is 2.75mm x 12mm, but the size on the delivery shaft is still 2.5mm x 12mm.Another quantum maverick balloon catheter was used to complete the procedure.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter within the shipping hoop and tyvek packaging and outer box.The outer box, labeled with lot batch was 27816227.The box was visually inspected and found unsealed with no other irregularity or issue.The tyvek product pouch, labeled with lot batch was 27816236 was visually inspected.The product pouch contained a unit still in the shipping hoop, for batch 27816227 and the seal was fully intact with no sign of tampering or attempt to open.Inspection of the remainder of the packaging and device presented no other damage or irregularities.Product analysis found there was a labeling issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14578132
Report Number2134265-2022-05578
Device Sequence Number5078090
Product Code LOX
UDI-Device Identifier08714729392651
UDI-Public08714729392651
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/10/2024
Device Model Number7010
Device Catalogue Number7010
Device Lot Number0027816227
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/10/2022
Supplement Date Received by Manufacturer08/12/2022
Initial Report FDA Received Date06/02/2022
Supplement Report FDA Received Date08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age42 YR
Patient SexFemale
Patient Weight60 KG
-
-