| Catalog Number |
303347 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date : unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Medical device manufacture date : unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that 1 bd syringe 5ml ll w/blunt plastic cannula had coring issues.The following information was provided by the initial reporter : the consumer facility reported issues with the blunt plastic cannulas that when the nursing staff are using these to draw up medications from a vial, they are pushing part of the rubber stopper into the bottle.Event date : unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H6.Investigation summary: bd syringes with blunt plastic cannula are specifically designed for use with preslit injection sites.They are not compatible with other injection sites.This is not the intended use of this product, hypodermic needles designed to puncture vials would provide a proper application.Since no samples displaying the reported condition were received a potential root cause could not be defined and corrective actions are not necessary.It is likely based on description that this product is not being used for its intended application.H3 other text : see h.10.
|
| |
|
Event or Problem Description
|
|
It was reported that 1 bd syringe 5ml ll w/blunt plastic cannula had coring issues.The following information was provided by the initial reporter : the consumer facility reported issues with the blunt plastic cannulas that when the nursing staff are using these to draw up medications from a vial, they are pushing part of the rubber stopper into the bottle.Event date : unknown.
|
| |
|