• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941812300
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the severely calcified vessel.A 3.00 x 12mm synergy xd drug-eluting stent was advanced for treatment.However, during the procedure, the proximal shaft broke while trying to advance through the calcified lesion.The device was completely removed and the procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
The synergy xd mr us 3.00 x 12mm stent delivery system (sds) was returned for analysis without the distal shaft containing the inner outer shaft polymer extrusion, stent, balloon, and tip.A visual and tactile examination of the hypotube shaft found multiple kinks along several locations of the hypotube shaft.A break at the wire exchange port was noted.The distal shaft polymer extrusion section of the device not returned so could not be reviewed.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the severely calcified vessel.A 3.00 x 12mm synergy xd drug-eluting stent was advanced for treatment.However, during the procedure, the proximal shaft broke while trying to advance through the calcified lesion.The device was completely removed and the procedure was completed with another of the same device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14579042
Report Number2134265-2022-06018
Device Sequence Number14712693
Product Code NIQ
UDI-Device Identifier08714729980957
UDI-Public08714729980957
Combination Product (Y/N)Y
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/08/2023
Device Model NumberH7493941812300
Device Lot Number0028398877
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer06/23/2022
Initial Report FDA Received Date06/02/2022
Supplement Report FDA Received Date06/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-