| Model Number |
357512 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A passeo-18 balloon catheter was chosen for treatment of 200mm long lesion in an artery of the lower extremity.During inflation the balloon could not be expanded as expected.The device was withdrawn, and a pinhole was detected.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned instrument was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the balloon has been inflated and was deflated in the as-returned state.Upon inflation the pressure could not be held.A fine jet of water was seen to emerge from the proximal balloon portion.Microscopic inspection showed a pinhole in the balloon at the proximal edge of the proximal x-ray marker.In close vicinity of the pinhole a long scratch was observed which have likely been caused by a hard, sharp-edged object such as e.G.Anatomical structure.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test and a pressure test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no material or manufacturing related root cause was determined.The root cause for the reported event is most likely related to the patients anatomy.
|
| |
|
Search Alerts/Recalls
|