| Model Number |
357511 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A passeo-18 balloon catheter was chosen for pre-dilatation of a moderately calcified lesion (95 percent stenosis degree) in the moderately tortuous sfa.During inflation up to 6 atm, it was noted that contrast medium was leaking out of the balloon.The balloon catheter was withdrawn and a pinhole was detected.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned instrument was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned product revealed that the balloon has burst transversally.The balloon shows clear signs of inflation.In close vicinity of the tear scratches were observed which have likely been caused by a hard, sharp-edged object such as e.G.Anatomical structure.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test and a pressure test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no material or manufacturing related root cause was determined.The root cause for the reported event is most likely related to the patients anatomy.
|
| |
|
Search Alerts/Recalls
|