• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURNEY FEM OX NP BCS RT SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURNEY FEM OX NP BCS RT SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED Back to Search Results
Model Number 74021113
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Limb Fracture (4518)
Date of Event 05/25/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a tka construct had been implanted on the patient¿s right knee on (b)(6) 2007, the patient experienced a periprosthetic multi-fragmentary rorabeck type ii fracture of the distal femoral spiral on the right limb due to a fall.To treat this adverse event, the patient underwent an open reduction and osteo-synthetic surgery on (b)(6) 2021.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
Corrected data: d10 (insert identity updated), h6 (health effect - clinical code).
 
Additional Manufacturer Narrative
(concomitant product added: patellar implant).
 
Additional Manufacturer Narrative
H10: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, based on the information provided, the reported periprosthetic fracture and subsequent open reduction and internal fixation procedure are related to the patient¿s fall and not the implant.The implant was not modified during the surgery and was noted to intact without loosening.The patient impact beyond that which has been reported cannot be determined.No further clinical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that fracture has been identified as possible adverse effects.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A potential cause that could contribute to the reported event include traumatic injury.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY FEM OX NP BCS RT SZ 3
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14579686
Report Number1020279-2022-02729
Device Sequence Number1996302
Product Code JWH
UDI-Device Identifier03596010554055
UDI-Public03596010554055
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K042515
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/28/2015
Device Model Number74021113
Device Catalogue Number74021113
Device Lot Number05LM06880
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2022
Supplement Date Received by Manufacturer07/11/2022
07/14/2022
07/29/2022
Initial Report FDA Received Date06/02/2022
Supplement Report FDA Received Date07/13/2022
07/18/2022
08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/30/2005
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
74022213; 74023231; PN: 74022213 / LOT: UNKNOWN; PN: 74023234 / LOT: 09AM20910; PN: 74024832 / LOT: 15EM14663
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age61 YR
Patient SexFemale
Patient Weight64 KG
-
-