Model Number M0035475200
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an endovascular procedure there was resistance when the subject coil was advanced through the middle of the microcatheter.The subject coil got stretched and prematurely detached after repeated repositioning.The procedure was completed successfully by replacing the device.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes) system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, visual testing as well as functional testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information provided by the customer was the device was prepared as per the dfu, that there was no damage noted to the packaging prior to opening and the device was confirmed to be in good condition during preparation/prior to use on the patient, and that continuous flush was set up and maintained during the procedure.The patients anatomy was very meandering, this may have contributed to the reported friction felt during the procedure.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned.H3 other text : the device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during an endovascular procedure there was resistance when the subject coil was advanced through the middle of the microcatheter.The subject coil got stretched and prematurely detached after repeated repositioning.The procedure was completed successfully by replacing the device.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 SOFT 5MM X 20CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14579986
Report Number3008881809-2022-00263
Device Sequence Number2057497
Product Code HCG
UDI-Device Identifier04546540676511
UDI-Public04546540676511
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/13/2024
Device Model NumberM0035475200
Device Catalogue NumberM0035475200
Device Lot Number23373783
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/12/2022
Supplement Date Received by Manufacturer07/20/2022
Initial Report FDA Received Date06/02/2022
Supplement Report FDA Received Date08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROCATHETER (UNKNOWN); UNKNOWN MICROCATHETER
Patient SexUnknown
-
-