| Catalog Number |
RBYPOD5 |
| Medical Device Problem Codes |
Physical Resistance/Sticking (4012); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the gastroduodenal artery (gda) using pod coils and a non-penumbra microcatheter.It should be noted that the patient''s anatomy was tortuous.During the procedure, the physician successfully implanted three non-penumbra coils into the distal target site.After advancing a pod coil into the proximal target site, the physician reported that the coil would not take the desired shape.Therefore, the pod coil was removed.After flushing the microcatheter, the pod coil was re-inserted and became stuck in the middle of the microcatheter.While attempting to move the pod coil, resistance was encountered and the pod coil unintentionally detached inside the microcatheter when the distal end of the coil was in the middle of the microcatheter.Therefore, the microcatheter containing the detached pod coil was removed from the patient.The procedure was completed using a non-penumbra coil and a new microcatheter.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned pod5 confirmed that the embolization coil was detached from the pusher assembly.Evaluation revealed that the pet lock was separated on the proximal end of the pusher assembly, and the pull wire was retracted out of the ddt.If this occurs, the embolization coil will detach from its pusher assembly.A demonstration coil was advanced through the non-penumbra microcatheter during evaluation without an issue.Therefore, the root cause of the returned pod coil getting stuck in the middle of the non-penumbra microcatheter during the procedure could not be determined.Penumbra products are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|
| |
|
Search Alerts/Recalls
|