An olympus representative reported to olympus technical assistance center (tac) on behalf of the customer, the sonicbeat 5 mm, 45 cm, front-actuated grip could not connect correctly to the cord during a therapeutic procedure.Subsequently, the procedure was completed.Upon inspection of the customer¿s returned device, it was observed, the grasping portion of the jaw was missing.This report is being submitted for the malfunction found during evaluation.There was no harm or user injury reported due to the event.
|
|
The device was returned to olympus for evaluation.The device was not returned in the original package.The device was attached to the usg-400/esg-400 and a probe check was performed; the device passed the probe check.Both switches were checked and found both switches were functional.A visual inspection was performed on the device and there was no tissue build up noted.The polytetrafluoroethylene (ptfe) pad (teflon pad) was missing due to the broken/missing grasping part, which was not returned for evaluation.The distal end of the device was inspected under a microscope and found no visual indication of damage to the probe unit; however, the grasping portion of the jaw is missing.The handle load was normal and the rotation of the knob torque was normal and smooth.Investigation activities have been opened to manage the actions related to this issue and any required mdr reporting.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The results did not confirm that the device was not properly connected to the cord, so a definitive root cause cannot be identified, the following information is included in the instructions for use (ifu): ¿do not activate output while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿when cutting or vessel sealing is performed, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ ¿during the treatment, do not activate output while applying the probe tip to the tissue with a strong force, grasping thick tissue or twisting the handle.Also, do not activate the output, while torque is applied to tissue over the handle, in this instance release the torque, re-grasp the tissue and re-activate output.Otherwise, the probe tip and/or grasping section may be damaged, which may result in falling of the probe tip and/or tissue pad.¿ olympus will continue to monitor the performance of this device.
|