Age, weight and ethnicity: unknown/asked but unavailable.If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).Attempts were made to obtain the missing information; however, the account did not have the information.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
|
Additional information: section d9: device available for evaluation? yes.Section d9: returned to manufacturer on: 6/7/2022.Section h3: device evaluated by manufacturer ¿ yes.Device evaluation: the complaint lens was received taped to the original lens case.Visual inspection under magnification revealed that the complaint lens was received cut in half (one half missing) and taped to the lens case.The lens was removed from the case and cleaned, revealing no further issues with the lens.Based on the return condition of the lens, no further product evaluation could be identified.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|