MEDACTA INTERNATIONAL SA GMK-PRIMARY TIBIAL INSERT PS FIXED SIZE 3/10MM; KNEE TIBIAL INSERT Back to Search Results
Model Number 02.07.0310PSF
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The insert screw broke while using a wrong screwdriver.It was decided to remove the cemented tibial tray and the insert and replace them with the same range of products in order to complete the surgery.The surgery was completed successfully.
 
Additional Manufacturer Narrative
Batch review performed on 13 may 2022: lot 2116761: (b)(4) items manufactured and released on 12-jan-2022.Expiration date: 2026-12-19.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
 
Additional Manufacturer Narrative
The item returned for analysis on september 20th 2022.Visual inspection performed by r&d knee manager: the fixation screw of the gmk primary ps tibial insert broke in the attempt of screwing into the tibial tray.As reported in the event description, the applied torqued has exceeded the maximum torque that the screw can withstand, causing the breakage.The gmk primary surgical technique requires the use of the 3.5nm torque limiter screwdriver to achieve the optimal locking of the screw.The issue was most likely caused by the use of the wrong screwdriver during the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-PRIMARY TIBIAL INSERT PS FIXED SIZE 3/10MM
Common Device Name
KNEE TIBIAL INSERT
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key14591430
Report Number3005180920-2022-00438
Device Sequence Number2021257
Product Code JWH
UDI-Device Identifier07630030817069
UDI-Public07630030817069
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K090988
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.07.0310PSF
Device Catalogue Number02.07.0310PSF
Device Lot Number2116761
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2022
Supplement Date Received by Manufacturer05/11/2022
Initial Report FDA Received Date06/03/2022
Supplement Report FDA Received Date10/10/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-