| Model Number |
10742326 |
| Medical Device Problem Codes |
Circuit Failure (1089); Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to siemens that a malfunction occurred while operating the somatom force ct system.The user reported that the phs (patient table) did not move to the scanning start position after a topo scan (low dose overview scan).As a result, the male pediatric patient, who is less than one year in age, was scanned again on another ct system.It was reported that the patient is in good health and we are unaware of any impact to the state of health of the patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.A defective sp-wire-sensor has been identified and replaced.No general product issue could be identified.This report has been submitted with an abundance of caution.
|
| |
|