| Model Number |
09P4120 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer observed falsely decreased alinity i sirolimus results on one patient specimen.The results provided were: on (b)(6) 2022.Sid (b)(6) =2.98 ng/l /repeated=8.05 ng/l /repeated few times=8.0, 8.1 and 8.7 ng/l which matched with patient¿s previous history.There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint evaluation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, device history record review, accuracy testing of alinity i sirolimus reagent lot 30297fp00.The ticket search determined that there is normal complaint activity for the likely cause lot.A review of complaints determined that there are no trends for the product for the complaint issue.Return testing was not completed as returns were not available.The ticket trending review of at least 12 months complaint data for the likely cause list number does not identify any adverse or non-statistical trend.Accuracy testing was performed with a retained kit of lot 30297fp00, and all specifications were met indicating that the lot is performing acceptably.Device history record review did not identify any non-conformances or deviations with the complaint lot and complaint issue.A review of the manufacturing documentation did not identify any issues associated with the customer¿s observation.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation no product deficiency was identified for the alinity i sirolimus reagent lot 30297fp00.
|
| |
|
Search Alerts/Recalls
|