| Model Number |
80617 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the coating of renegade was fractured.The target lesion was located in the liver.A renegade stc 18 catheter-infusion was selected for use.During preparation, it was noted that the coating of renegade was fractured.The procedure was completed with another of same device.No patient complications were reported and patient was stable post procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that the coating of renegade was fractured.The target lesion was located in the liver.A renegade stc 18 catheter-infusion was selected for use.During preparation, it was noted that the coating of renegade was fractured.The procedure was completed with another of same device.No patient complications were reported and patient was stable post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was returned for evaluation.The device was inspected for any damage or irregularities.The renegade stc 18 was received in the shipping hoop.The hoop was hydrated, and the device was removed.No fractures or damage to the coating was visible to the eye or with the microscope.Tactile analysis could find no places where there was a problem with the coating.It was found to be slippery where it should be when hydrated.An in house guide wire was inserted and the renegade stc 18 was advanced through an in house terumo guide catheter without resistance.No other issues were identified during the product analysis.The reported issue was not confirmed.
|
| |
|
Search Alerts/Recalls
|