Model Number 80617
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the coating of renegade was fractured.The target lesion was located in the liver.A renegade stc 18 catheter-infusion was selected for use.During preparation, it was noted that the coating of renegade was fractured.The procedure was completed with another of same device.No patient complications were reported and patient was stable post procedure.
 
Event or Problem Description
It was reported that the coating of renegade was fractured.The target lesion was located in the liver.A renegade stc 18 catheter-infusion was selected for use.During preparation, it was noted that the coating of renegade was fractured.The procedure was completed with another of same device.No patient complications were reported and patient was stable post procedure.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the device was returned for evaluation.The device was inspected for any damage or irregularities.The renegade stc 18 was received in the shipping hoop.The hoop was hydrated, and the device was removed.No fractures or damage to the coating was visible to the eye or with the microscope.Tactile analysis could find no places where there was a problem with the coating.It was found to be slippery where it should be when hydrated.An in house guide wire was inserted and the renegade stc 18 was advanced through an in house terumo guide catheter without resistance.No other issues were identified during the product analysis.The reported issue was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENEGADE STC 18
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14592571
Report Number2134265-2022-06016
Device Sequence Number5517185
Product Code DQO
UDI-Device Identifier08714729444442
UDI-Public08714729444442
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K023681
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/24/2024
Device Model Number80617
Device Catalogue Number80617
Device Lot Number0027895079
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/13/2022
Supplement Date Received by Manufacturer07/25/2022
Initial Report FDA Received Date06/03/2022
Supplement Report FDA Received Date08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexMale
Patient Weight75 KG
-
-