• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ZENITH FLEX WITH SPIRAL-Z TECHNOLOGY AAA ENDOVASCULAR GRAFT ILIAC LEG; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ZENITH FLEX WITH SPIRAL-Z TECHNOLOGY AAA ENDOVASCULAR GRAFT ILIAC LEG; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number N/A
Medical Device Problem Codes Partial Blockage (1065); Fluid/Blood Leak (1250)
Health Effect - Clinical Codes Failure of Implant (1924); Thrombosis/Thrombus (4440)
Date of Event 05/03/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a right zenith flex with spiral-z technology aaa endovascular graft iliac leg occluded following implantation on (b)(6) 2017.During the index procedure on (b)(6) 2017, the patient received a main body device and two iliac legs.On (b)(6) 2022, the patient presented with renal colic.A ct scan revealed an occluded right zenith iliac leg graft and a type 1a endoleak on the main body graft.An angiogram was performed and the graft was relined.Additional information regarding the event and patient outcome has been requested but is currently unavailable.
 
Additional Manufacturer Narrative
The zenith main body graft that experienced a type 1a endoleak is not cleared for sale in the u.S.This device is not same or similar to a device sold in the u.S.It should be noted that zsle- devices are distributed via one-device lots.(b)(6).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged or unavailable.H6: annex e: e2121, endoleak.Additional information: b5, d10, h6 - annex a investigation ¿ evaluation.(b)(6) hospital sa informed cook on (b)(6) 2022 of a (b)(6) patient with occlusion of the right iliac limb.The complaint device was zenith flex with spiral-z technology aaa endovascular graft iliac leg (rpn: zsle-13-74-zt, lot: 7574415).The initial procedure was performed on (b)(6) 2017, in which the following devices were implanted zalb-32-84 (lot: 7490577), zsle-13-74-zt (lot: 7574415) and zsle-20-74-zt (lot: 7462325).The physician performed an additional procedure relining the graft.There have been no reports on the outcome of the relining procedure.A type 1b endoleak in the right zsle-13-74-zt was discovered during the investigation for the right limb occlusion.Reviews of the documentation, including the complaint history, device history record, drawing, instructions for use (ifu), manufacturing instructions, quality control procedures and specifications of the device, were conducted during the investigation.The complaint device was not returned for evaluation; therefore, no physical examination could be performed.However, medical imaging was provided by the facility and was reviewed by an expert physician.Thrombosis of the right zsle-13-74 was confirmed.External compression, acuity angulation, severe tortuosity and outflow limitation were absent.The stent shortening was not subjectively excessive and likely necessary to accommodate the right common iliac artery curvature.Stent implantation into the main body superior the flow divider may have contributed to thrombosis by creating a thrombogenic blind end in the distal main body.Right common iliac dilation beyond the right zsle nominal diameter represented a type 1b endoleak.The seal zone fusiform dilation and right common iliac artery dilation were likely related to intrinsic aortic disease.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that appropriate inspections are in place relative to the reported failure.A review of the device history record (dhr) for lot 7574415, graft lot sa7318842 and graft kit sa7064975 found no nonconformances that could have contributed to the reported failure mode.The graft subassembly lot (1649922.38) was sewn by approved vendor sei medical s.A.Upon incoming inspection zero nonconformance¿s were detected.It should be noted that there were two other complaints associated with the final product lot number; however, these represent different unrelated events in the same patient.Cook also reviewed product labeling.The device was packaged with ifu t_zaaasz_rev3.The ifu includes the following, warnings, precautions, and instructions for proper placement of device.4 warnings and precautions 4.1 general additional endovascular interventions or conversion to standard open surgical repair following initial endovascular repair should be considered for patients experiencing enlarging aneurysms, unacceptable decrease in fixation length (vessel and component overlap) and/or endoleak.An increase in aneurysm size and/or persistent endoleak or migration may lead to aneurysm rupture.Patients experiencing reduced blood flow through the graft limb and/or leaks may be required to undergo secondary interventions or surgical procedures.4.2 patient selection, treatment and follow-up adequate iliac or femoral access is required to introduce the device into the vasculature.Access vessel diameter (measured inner wall to inner wall) and morphology (minimal tortuosity, occlusive disease and/or calcification) should be compatible with vascular access techniques and delivery systems of a 14 french to 16 french vascular access techniques and delivery systems of a 14 french to 16 french vascular introducer sheath.Vessels that are significantly calcified, occlusive, tortuous, or thrombus-lined may preclude placement of the endovascular graft and/or may increase the risk of embolization.A vascular conduit technique may be necessary to achieve in some patients.Pre-existing regions of stenosis/narrowing (less than approximately 20 mm id in aorta or 7 to 8 mm id in the iliacs) have been shown to increase the risk of a thromboembolic event (e.G., graft limb occlusion).The potential for this increased risk in these patients may preclude placement of an endovascular graft.Dilatation of these regions with a noncompliant balloon and/or stent placement may be necessary to help assure maintained graft patency and to reduce the risk of a thromboembolic event.Additionally, the completion angiogram (with stiff wire guides removed) should be reviewed carefully to determine if further treatment in these regions is necessary (e.G., adjunctive ballooning or stenting).Failure to remove stiff wire guide prior to the angiogram could mask any limb kinking or narrowing that might occur when the wire guide is removed.Follow-up imaging should be carefully reviewed for narrowing within the graft leg.Patients with a graft leg lumen of less than approximately 5 mm id may be at increased risk of a thromboembolic event (e.G., graft limb occlusion).Reintervention (e.G., noncompliant ballooning or stenting in these regions) should be considered to help assure maintained graft patency and to reduce the risk of a thromboembolic event.Patients with poor outflow or a hypercoagulable state (e.G., cancer) may be at an increased risk of a thromboembolic event.The zenith spiral-z aaa iliac leg with the z-trak introduction system is not recommended in patients who cannot tolerate contrast agents necessary for intraoperative and postoperative follow-up imaging.All patients should monitored closely and checked periodically for a change in the condition of their disease and the integrity of the endoprosthesis.Successful patient selection requires specific imaging and accurate measurements; please see section 4.3, pre-procedure measurement techniques, and imaging.4.3 pre-procedure measurement technique and imaging clinical experience indicates that contrast-enhanced spiral computed tomographic angiography (cta) with 3-d reconstruction is the strongly recommended imaging modality to accurately assess patient anatomy prior to treatment with the zenith spiral-z aaa iliac leg.If contrast-enhanced spiral cta with 3-d reconstruction is not available, the patient should be referred to a facility with these capabilities.Lengths all patients should be advised that endovascular treatment requires life-long, regular follow-up to assess their health and the performance of their endovascular graft.Patients with specific clinical findings (e.G., endoleaks, enlarging aneurysm or changes in the structure or position of the endovascular graft) should receive enhanced follow-up.Specific follow-up guidelines are described in section 12, imaging guidelines and postoperative follow-up.After endovascular graft placement, patients should be regularly monitored for perigraft flow, aneurysm growth or changes in the structure or position of the endovascular graft.At a minimum, annual imaging is required, including: 1) abdominal radiographs to examine device integrity (separation between components or stent fracture) and 2) contrast and non-contrast ct to examine aneurysm changes, perigraft flow, patency, tortuosity, and progressive disease.If renal complications or other factors preclude the use of image contrast media, abdominal radiographs and duplex ultrasound may provide similar information.4.5 implant procedure do not continue advancing any portion of the delivery system if resistance is felt during advancement of the wire guide or delivery system.Stop and assess the cause of resistance; vessel, catheter or graft damage may occur.Exercise particular care in areas of stenosis, intravascular thrombosis, or in calcified or tortuous vessels.Fluoroscopy should be used during introduction and deployment to confirm proper operation of the delivery system components, proper placement of the graft, and desired procedural outcome.Excessive overlap 10 mm above the main body bifurcation may increase the risk of limb thrombosis.Inaccurate placement and/or incomplete sealing of the zenith spiral-z aaa iliac leg within the vessel may result in increased risk of endoleak, migration or inadvertent occlusion of the internal iliac arteries.Avoid damaging the graft or disturbing graft positioning after placement in the event re-instrumentation (secondary intervention) of the graft is necessary.5 adverse events 5.2 potential adverse events claudication (e.G., buttock, lower limb) endoprosthesis: improper component placement; incomplete component deployment; component migration; component separation from another graft component; suture break; occlusion; infection; stent fracture; graft material wear; dilatation; erosion; puncture; perigraft flow; and corrosion graft or native vessel occlusion aneurysm enlargement claudication (e.G., buttock, lower limb) endoleak 7 patient selection and treatment co-morbidities (e.G., cardiac, pulmonary, or renal insufficiency prior to surgery, morbid obesity) patient¿s anatomical suitability for endovascular repair patient¿s ability to tolerate general, regional, or local anesthesia iliofemoral access vessel size and morphology (minimal thrombus, calcification and/or tortuosity) should be compatible with vascular access techniques and accessories of the delivery profile of a 14 french to 16 french vascular introducer sheath freedom from significant femoral/iliac artery occlusive disease that would impede flow through the endovascular graft 8 patient counseling information all patients should be advised that endovascular treatment requires life-long, regular follow-up to assess their health and the performance of their endovascular graft.Patients with specific clinical findings (e.G., endoleaks, enlarging aneurysms or changes in the structure or position of the endovascular graft) should receive enhanced follow-up.Specific follow-up guidelines are described in section 12, imaging guidelines and postoperative follow-up.Patients should be counseled on the importance of adhering to the follow-up schedule, both during the first year and at yearly intervals thereafter.Patients should be told that regular and consistent follow-up is a critical part of ensuring the ongoing safety and effectiveness of endovascular treatment of aaas.At a minimum, annual imaging and adherence to routine postoperative follow-up requirements is required and should be considered a life-long commitment to the patient¿s health and well-being.Physicians must advise all patients that it is important to seek prompt medical attention if they experience signs of limb occlusion, aneurysm enlargement or rupture.Signs of graft limb occlusion include pain in the hip(s) or leg(s) during walking or at rest or discoloration or coolness of the leg.Aneurysm rupture may be asymptomatic, but usually presents as: pain; numbness; weakness in the legs; any back, chest, abdominal or groin pain; dizziness; fainting; rapid heartbeat or sudden weakness.Physicians should refer patients to the patient guide regarding risks occurring during or after implantation of the device.Procedure-related risks include cardiac, pulmonary, neurologic, bowel and bleeding complications.Device-related risks include occlusion, endoleak, aneurysm enlargement, fracture, potential for reintervention and open surgical conversion, rupture, and death (see section 5.1, observed adverse events and section 5.2, potential adverse events).The physician should complete the patient i.D.Card and give it to the patient so that he/she can carry it with him/her at all times.The patient should refer to the card anytime he/she visits additional health practitioners, particularly for any additional diagnostic procedures (e.G., mri).12 imaging guidelines and postoperative follow-up 12.1 general all patients should be advised that endovascular treatment requires lifelong, regular follow-up to assess their health and the performance of their endovascular graft.Patients with specific clinical findings (e.G., endoleaks, enlarging aneurysms or changes in the structure or position of the endovascular graft) should receive additional follow-up.Patients should be counseled on the importance of adhering to the follow-up schedule, both during the first year and at yearly intervals thereafter.Patients should be told that regular and consistent follow-up is a critical part of ensuring the ongoing safety and effectiveness of endovascular treatment of aaas.Physicians should evaluate patients on an individual basis and prescribe follow-up relative to the needs and circumstances of each individual patient.The minimum requirement for patient follow-up (described in the instructions for use for the zenith aaa device that was used) should be maintained even in the absence of clinical symptoms (e.G., pain, numbness, weakness).Patients with specific clinical findings (e.G., endoleaks, enlarging aneurysms or changes in the structure or position of the stent graft) should receive follow-up at more frequent intervals.The combination of contrast and non-contrast ct imaging provides information on aneurysm diameter change, endoleak, patency, tortuosity, progressive disease, fixation length and other morphological changes.Duplex ultrasound imaging may provide information on aneurysm diameter change, endoleak, patency, tortuosity, and progressive disease.In this circumstance, a non-contrast ct should be performed to use in conjunction with the ultrasound.Ultrasound may be a less reliable and sensitive diagnostic method compared to ct.After review of the ifu, cook has concluded the device labeling contains appropriate warnings, precautions and instructions to the user related to the reported failure.Evidence provided by the complaint facility, device failure analysis, dhr, complaint history, manufacturing documents, and verification testing, suggests that the device was manufactured to specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.Based on the information provided, expert review of medical imaging and the results of our investigation, a definitive cause for the right limb occlusion could not be established.It is possible that device placement may have contributed to this incident.The image reviewer noted the leg was implanted 12.2mm superior the flow divider.The left zsle was implanted 6.1mm superior the flow divider.The ifu states, ¿excessive overlap 10 mm above the main body bifurcation may increase the risk of limb thrombosis¿.The complaint is confirmed based on customer testimony and the provided imaging.A definitive cause for the type 1b endoleak could also not be established.It is possible that disease progression/patient anatomy may have contributed to this incident.The image reviewer noted the right common iliac artery had dilated beyond the zsle.This permitted a type ib endoleak to fill the aneurysm sac.The appropriate personnel have been notified.Per the risk assessment no further action is required.We will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
A type 1b endoleak in the right zsle-13-74-zt was identified upon expert image review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZENITH FLEX WITH SPIRAL-Z TECHNOLOGY AAA ENDOVASCULAR GRAFT ILIAC LEG
Common Device Name
MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14593690
Report Number1820334-2022-00998
Device Sequence Number2057950
Product Code MIH
UDI-Device Identifier10827002552323
UDI-Public(01)10827002552323(17)200117(10)7574415
Combination Product (Y/N)N
PMA/510(K) Number
P020018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/17/2020
Device Model NumberN/A
Device Catalogue NumberZSLE-13-74-ZT
Device Lot Number7574415
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/13/2022
Supplement Date Received by Manufacturer08/17/2022
Initial Report FDA Received Date06/03/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/17/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
COOK ILIAC LEG GRAFT RPN:ZSLE-20-74-ZT, LOT7462325; COOK MAIN BODY GRAFT, RPN: ZALB-32-84, LOT 7490577; ZALB-32-84; ZSLE-20-74-ZT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexMale
-
-