• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK SOLUTION ADMINISTRATION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK SOLUTION ADMINISTRATION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8857
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the clamp on a vented paclitaxel set failed to prevent fluid/drops from falling from the fluid container when fully engaged.The issue was identified during priming.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
H10: the device was received for evaluation.Visual inspection using the naked eye did not identify any abnormalities that could have contributed to the reported condition.All components were correctly placed and according to specifications.Functional testing was performed including clear passage and pressure testing; and the results were satisfactory.Additional testing was performed including priming the set with a bag of sterile water and by closing the slide clamp in 10 seconds and no defects were observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK SOLUTION ADMINISTRATION SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14593872
Report Number1416980-2022-02878
Device Sequence Number2095994
Product Code FPA
UDI-Device Identifier00085412046372
UDI-Public(01)00085412046372
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K160007
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/01/2026
Device Catalogue Number2C8857
Device Lot NumberR21F30082
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/10/2022
Supplement Date Received by Manufacturer07/04/2022
Initial Report FDA Received Date06/03/2022
Supplement Report FDA Received Date07/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-