| Catalog Number |
2C8857 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the clamp of a vented paclitaxel set failed to prevent fluid/drops from falling from the fluid container when fully engaged.The issue was identified during priming.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to d4 (expiration date), d9, h3, h4, h6 and h10.H4: device manufactured between may 21, 2018 to may 22, 2018.H10: the device was received for evaluation.Visual inspection using the naked eye did not identify any abnormalities that could have contributed to the reported condition.All components were correctly placed and according to specifications.Functional testing was performed including clear passage and pressure testing; and the results were satisfactory.Additional testing was performed including priming the set with a bag of sterile water and by closing the slide in 10 seconds and no defects were observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|