During the processing of this complaint, attempts were made to obtain complete event, patient, and device information.It was reported that during the implantation of a ct lucia 602 lens it was noted that the lens was damaged.Thus, the lens was removed, and a spare lens used to successfully complete the procedure.There was no patient injury or clinically significant delay in procedure reported.It was reported the customer is using an emerald series injector to inject the lens.No additional information provided.The device was discarded thus, a proper device analysis could not be completed nor the reported issue confirmed.It was stated by the customer that there was no damage noted to the device during preparation for use.Additionally, it was also stated that the customer is using an emerald series to implant the lenses.The device history records were reviewed and there were no non-conformities or deviations noted during the manufacturing of the lenses that would have contributed to the nature of this complaint.Additionally, lenses are 100% inspected before they leave our manufacturing site.Therefore, we are confident that the lens was processed per standard operation procedures and inspections and met all of the criteria for release.Additionally, there was no damage noted to the lens during preparation for use which suggest a product deficiency did not contribute to the reported issue.Damaged haptics are known to be caused by numerous factors including, but not limited to: loading strategy; lens placement technique; poor handling during folding and inserting; patient pathology.We have reviewed the information provided and at this time there is no indication of a product quality issue with respect to the manufacturing, design, or labeling of the device.It appears that the reported issue is linked to user error.However, the need to explant the lens is considered need for medical intervention in order to prevent and avoid permanent impairment or damage to a body structure.
|
|
It was reported that during the implantation of a ct lucia 602 lens it was noted that the lens was damaged.Thus, the lens was removed, and a spare lens used to successfully complete the procedure.There was no patient injury or clinically significant delay in procedure reported.It was reported the customer is using an emerald series injector to inject the lens.No additional information provided.
|