• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARL ZEISS MEDITEC PRODUCTION LLC CT LUCIA 602

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARL ZEISS MEDITEC PRODUCTION LLC CT LUCIA 602 Back to Search Results
Model Number 003500-0050-680
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Visual Impairment (2138)
Date of Event 05/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
During the processing of this complaint, attempts were made to obtain complete event, patient, and device information.It was reported that during the implantation of a ct lucia 602 lens it was noted that the lens was damaged.Thus, the lens was removed, and a spare lens used to successfully complete the procedure.There was no patient injury or clinically significant delay in procedure reported.It was reported the customer is using an emerald series injector to inject the lens.No additional information provided.The device was discarded thus, a proper device analysis could not be completed nor the reported issue confirmed.It was stated by the customer that there was no damage noted to the device during preparation for use.Additionally, it was also stated that the customer is using an emerald series to implant the lenses the device history records were reviewed and there were no non-conformities or deviations noted during the manufacturing of the lenses that would have contributed to the nature of this complaint.Additionally, lenses are 100% inspected before they leave our manufacturing site.Therefore, we are confident that the lens was processed per standard operation procedures and inspections and met all of the criteria for release.Additionally, there was no damage noted to the lens during preparation for use which suggest a product deficiency did not contribute to the reported issue.Damaged haptics are known to be caused by numerous factors including, but not limited to: loading strategy; lens placement technique; poor handling during folding and inserting; patient pathology.We have reviewed the information provided and at this time there is no indication of a product quality issue with respect to the manufacturing, design, or labeling of the device.It appears that the reported issue is linked to user error.However, the need to explant the lens is considered need for medical intervention in order to prevent and avoid permanent impairment or damage to a body structure.
 
Event or Problem Description
It was reported that during the implantation of a ct lucia 602 lens it was noted that the lens was damaged.Thus, the lens was removed, and a spare lens used to successfully complete the procedure.There was no patient injury or clinically significant delay in procedure reported.It was reported the customer is using an emerald series injector to inject the lens.No additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CT LUCIA 602
Common Device Name
CT LUCIA 602
Manufacturer (Section D)
CARL ZEISS MEDITEC PRODUCTION LLC
1040 s vintage ave. bld. a
ontario CA 91761
Manufacturer (Section G)
CARL ZEISS MEDITEC PRODUCTION LLC
1040 s vintage ave. bld. a
ontario CA 91761
Manufacturer Contact
charlie ortiz
1040 s vintage ave. bld. a
ontario, CA 91761
9099065165
MDR Report Key14594296
Report Number3010126268-2022-00014
Device Sequence Number14905656
Product Code HQL
UDI-Device Identifier00843045101017
UDI-Public(01)00843045101017(17)250930
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P100016-S008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/25/2025
Device Model Number003500-0050-680
Device Catalogue Number003500-0050-680
Device Lot Number3S201339
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/10/2022
Initial Report FDA Received Date06/03/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-