• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPINE WAVE, INC. PROFICIENT POSTERIOR CERVICAL SPINE SYSTEM; POSTERIOR CERVICO-THORACIC FIXATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPINE WAVE, INC. PROFICIENT POSTERIOR CERVICAL SPINE SYSTEM; POSTERIOR CERVICO-THORACIC FIXATION SYSTEM Back to Search Results
Model Number 11-8999
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
After approximately 5 months post-operatively, a surgical procedure was performed to replace locking screws that had dissociated from the screw heads in a multi-level cervico-thoracic construct.
 
Additional Manufacturer Narrative
The explanted locking screws were returned for evaluation.Post-operative radiographic images were received confirming the condition reported.Visual examination found wear and deformation consistent with the reported condition and use in a pedicle screw construct.Further evaluation found no functional defects and the production records did not identify any non-conformances to specification.Based on the results of the investigation, a definitive conclusion could not be drawn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFICIENT POSTERIOR CERVICAL SPINE SYSTEM
Common Device Name
POSTERIOR CERVICO-THORACIC FIXATION SYSTEM
Manufacturer (Section D)
SPINE WAVE, INC.
3 enterprise drive
suite 210
shelton CT 06484
Manufacturer (Section G)
SPINE WAVE, INC.
3 enterprise drive
suite 210
shelton CT 06484
Manufacturer Contact
ronald smith
3 enterprise drive
suite 210
shelton, CT 06484
2039449494
MDR Report Key14594654
Report Number3004638600-2022-00003
Device Sequence Number11277204
Product Code NKG
UDI-Device Identifier10840642111217
UDI-Public10840642111217
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K162639
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number11-8999
Device Catalogue Number11-8999
Device Lot Number662W220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2022
Initial Report FDA Received Date06/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-