• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number ZA9003
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a defect was observed on the intraocular lens (iol) upon insertion in the patient¿s left eye.Account indicated that the iol was removed and replaced in the same procedure.The replacement iol was of the same model and diopter power.No patient injury was noted and no medical intervention was required.No further information was provided.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation? yes.Section d9: returned to manufacturer: yes.Section d9: date returned to manufacturer: 7/8/2022.Section h3: device returned to manufacturer: yes.Device evaluation: visual inspection of the returned device under magnification revealed that the complaint lens was received crushed inside of the lens module and cut in half (one half missing).The lens was cleaned and, no lens damage could be identified.However, the haptic could be observed to be partially detached with the remaining portion of the haptic bent.Based on the return condition of the lens, no further product evaluation could be identified.The reported complaint issue was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14594805
Report Number3012236936-2022-01429
Device Sequence Number20216289
Product Code HQL
UDI-Device Identifier05050474528970
UDI-Public(01)05050474528970(17)250331
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model NumberZA9003
Device Catalogue NumberZA90030180
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date06/03/2022
Supplement Report FDA Received Date08/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/31/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EMERALDC CARTRIDGE, LOT CK05894.
Patient Age61 YR
Patient SexMale
-
-