| Model Number |
6290 |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the communicator power cord melted, and the communicator was not powering up.The patient reported a burning smell.A boston scientific company representative discussed with the patient and opted to replace the entire system.A return label was sent.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the communicator power cord melted, and the communicator was not powering up.The patient reported a burning smell.A boston scientific company representative discussed with the patient and opted to replace the entire system.No adverse patient effects were reported.The communicator was deactivated.A return label was sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory additional information from product analysis indicates that the allegation was confirmed.During visual analysis, the communicator adapter had a missing piece, and the power adapter was not snug when plugged in.The communicator was taking apart, and areas were burnt and could smell a smoky, fire smell.Resistance testing on a component was not performed due to the component was being completely burnt.The communicator behavior was consistent with a high-power electrical overstress event.If information is provided in the future, a supplemental report will be issued.
|
| |
|