• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number TN DXH 600 HEMATOLOGY SYSTEM
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient information is not applicable.There was no impact to patients as a result of this event.One technician had a drop of retic clear reagent flick into his eye while replacing the retic pack.The technician flushed his eye with water, stating there was no irritation or vision issue.Per lab exposure protocol, went to the emergency department (ed) to have his eye checked.The ed noted no real (eye) damage and prescribed the lab tech an antibiotic.The failure mode was use error; the technician was not wearing eye protection at time of incident.The dxh series, ifu, pn: b26647ag, states the following safety notice: beckman coulter, inc.Urges its customers to comply with all national health and safety standards such as the use of barrier protection.This may include, but is not limited to, protective eyewear, gloves, and suitable laboratory attire when operating or maintaining this or any other automated laboratory analyzer.Bec internal identifier: (b)(4).
 
Event or Problem Description
The customer reported while replacing the retic pack on their dxh 600 hematology instrument, the pick up tube flicked a drop of retic clear reagent into the technician¿s right eye.The technician was wearing lab coat and gloves (no face protection) at time of exposure.The technician flushed his eye with water, stating there was no irritation or vision issue.Per lab exposure protocol, went to the emergency department (ed) to have his eye checked.The ed noted no real (eye) damage and prescribed the lab tech an antibiotic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEM
Common Device Name
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196 2031
Manufacturer Contact
angela vettel
250 s. kramer blvd
brea, CA 92821
7149613625
MDR Report Key14596064
Report Number1061932-2022-00034
Device Sequence Number17531654
Product Code GKZ
UDI-Device Identifier15099590192242
UDI-Public(01)15099590192242(11)NO-DATA
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K140911
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTN DXH 600 HEMATOLOGY SYSTEM
Device Catalogue NumberB23858
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/14/2022
Initial Report FDA Received Date06/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-