• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US APG+ STRAIGHT HANDLE OUTER; EXTREMITY INSTRUMENTS : SCREWDRIVERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US APG+ STRAIGHT HANDLE OUTER; EXTREMITY INSTRUMENTS : SCREWDRIVERS Back to Search Results
Model Number 2230-00-025
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the handle will not stay locked into drill guide, the guide falls right off.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary:visual examination of the returned device found expected wear from the handle of the device, and functional testing found the device engaging with its mating devices as intended.Due to investigation findings the reported allegation cannot be confirmed.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APG+ STRAIGHT HANDLE OUTER
Common Device Name
EXTREMITY INSTRUMENTS : SCREWDRIVERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14596452
Report Number1818910-2022-10082
Device Sequence Number2021415
Product Code HXC
UDI-Device Identifier10603295102984
UDI-Public10603295102984
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2230-00-025
Device Catalogue Number223000025
Device Lot NumberPC3048820
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer07/11/2022
Initial Report FDA Received Date06/03/2022
Supplement Report FDA Received Date07/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/07/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-