Model Number 347.985
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2022, during inspection, the 2 broken small depth gauges and 2 broken screw forceps were found.There was no patient involvement.This report involves one (1) screw and plate holding forceps.This is report 3 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: the subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d10 h3, h4, h6: part: 347.985 lot: a7qa26 manufacturing site: tuttlingen release to warehouse date: 28 june 2007 a review of the device history records was performed for the finished device lot number, and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.¿ visual analysis of the returned sample revealed that the screw and plate holding forceps has a broken tips.The event occurred outside of the operating room.No other problems was reported.A dimensional inspection was not performed for the screw and plate holding forceps due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the screw and plate holding forceps would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? current revision was reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW AND PLATE HOLDING FORCEPS
Common Device Name
FORCEPS
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen CO 78532
GM   78532
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14596471
Report Number2939274-2022-02061
Device Sequence Number14012977
Product Code HTD
UDI-Device Identifier10886982193629
UDI-Public10886982193629
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number347.985
Device Catalogue Number347.985
Device Lot NumberA7QA26
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/05/2022
Supplement Date Received by Manufacturer06/13/2022
Initial Report FDA Received Date06/03/2022
Supplement Report FDA Received Date07/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; SCREW AND PLATE HOLDING FORCEPS
Patient SexUnknown
-
-