| Model Number |
VASOVIEWHEMPRO (OUS) VH-3000 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Trackwise id (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.Device discarded.
|
| |
|
Event or Problem Description
|
|
The hospital reported that during an endoscopic vein harvesting procedure, vasoview hemopro (ous) vh-3000 coagulation forceps appeared to be still working without the trigger switch, which made it impossible to use normally.The electrocoagulation forceps will continue to work without a trigger switch, and can only stop working when the power is disconnected.After that, the same product was replaced to complete the operation without causing any harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Trackwise # (b)(4).The device was not returned to maquet cardiac surgery for investigation, therefore no evaluation could be performed, however a photograph was provided.A photographic inspection was conducted.The jaws were observed to be opened and glowing red during a procedure.The toggle was not observed in the photograph.Based on the photographic inspection, the reported failure "electrical issue/self-activation" was confirmed.The lot history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|
| |
|
Search Alerts/Recalls
|