MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEWHEMPRO (OUS) VH-3000; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEWHEMPRO (OUS) VH-3000; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VASOVIEWHEMPRO (OUS) VH-3000
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Trackwise id (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.Device discarded.
 
Event or Problem Description
The hospital reported that during an endoscopic vein harvesting procedure, vasoview hemopro (ous) vh-3000 coagulation forceps appeared to be still working without the trigger switch, which made it impossible to use normally.The electrocoagulation forceps will continue to work without a trigger switch, and can only stop working when the power is disconnected.After that, the same product was replaced to complete the operation without causing any harm to the patient.
 
Additional Manufacturer Narrative
Trackwise # (b)(4).The device was not returned to maquet cardiac surgery for investigation, therefore no evaluation could be performed, however a photograph was provided.A photographic inspection was conducted.The jaws were observed to be opened and glowing red during a procedure.The toggle was not observed in the photograph.Based on the photographic inspection, the reported failure "electrical issue/self-activation" was confirmed.The lot history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEWHEMPRO (OUS) VH-3000
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key14596816
Report Number2242352-2022-00491
Device Sequence Number13977565
Product Code GEI
UDI-Device Identifier00607567700413
UDI-Public00607567700413
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K052274
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/13/2022
Device Model NumberVASOVIEWHEMPRO (OUS) VH-3000
Device Catalogue NumberVH-3000-W
Device Lot Number25160721
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2022
Supplement Date Received by Manufacturer07/26/2022
Initial Report FDA Received Date06/03/2022
Supplement Report FDA Received Date07/26/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
Patient Weight73 KG
-
-