| Model Number |
72204042 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an acl procedure, when starting the preparation of the femoral tunnel, the surgeon identified that the trunav retrograde drill was not progressing, then, upon manually rotating the device, the drill did not move, so he removed it and saw that the drill had broken outside.The procedure was completed without surgical delay using a s+n back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an acl procedure, when starting the preparation of the femoral tunnel, the surgeon identified that the trunav retrograde drill was not progressing, then, upon manually rotating the device, the drill did not move, so he removed it and saw that the drill had broken outside, no pieces fell inside patient.Patient¿s bone quality was good.The procedure was completed without surgical delay using a s+n back-up device.No further complications were reported.Patient¿s health condition is stable.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.It was determined the device contributed to the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the customer provided images finds an x-ray image of a joint with no metal fragments; a retrograde drill with both the inner drill shaft and the distal end of the device bent and a retrograde drill with a part of the internal shaft broken lose from the device.Per case details, the drill broke outside the patient.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The orange bezel at the proximal end is sheared away, the actuator wire and the retrograde drill are connected to the grey slider.There is debris on the device.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states two photos of the device were provided for review and confirms the reported breakage.The foreign body seen in the provided intraoperative photo is not believed to be a part of the trunav retrograde drill as the reporter confirmed the break was outside of the patient and no parts within the patient.The object in the fluoroscopic images is presumed to be unrelated to the reported event.Based on the information provided no clinical factors were found which would have contributed to the reported events.The root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
| |
|